研究目的
To study the possibility to use the Cherenkov light for the e?cient detection of 511 keV photons with the goal to use it in TOF-PET.
研究成果
The study demonstrated the possibility to build a Cherenkov PET detection module with high efficiency and good timing performance using PbF2 crystal and commercial MCP-PMT. The measured detection efficiency of 24% for 511 keV photons in a 10 mm thick crystal and a coincidence resolving time of 280 ps are suitable for use in PET. The time resolution is limited by the low number of detected optical photons, the PMTs performances, and the implemented readout scheme.
研究不足
The main factors limiting the time resolution of the large-surface detection module include the dispersion of the photon trajectories and the PMT timing resolution. Solutions to improve these limitations are proposed for future projects.