研究目的
To describe the new 10V PJVS system created at VNIIM and present results of its comparison with the conventional VNIIM JVS system in the range from 1 mV to 10 V.
研究成果
The comparison between the PJVS and JVS systems shows agreement at the subnanovolt level. The PJVS allows calibration of sampling systems at different modes with a μV/V accuracy.
研究不足
The study focuses on the comparison between two quantum voltage standards architectures and does not explore other potential applications or limitations of the PJVS system.
1:Experimental Design and Method Selection:
The study involved the development of a PJVS system and its comparison with a conventional JVS system.
2:Sample Selection and Data Sources:
The comparison was conducted using a 10 V PJVS system and a conventional dc JVS system.
3:List of Experimental Equipment and Materials:
The PJVS equipment includes a cryogenic probe, bias source, and microwave source.
4:Experimental Procedures and Operational Workflow:
The comparison layout and numerical results are provided, detailing the setup and measurements.
5:Data Analysis Methods:
The results of the comparison are analyzed to show agreement at the subnanovolt level.