To characterize the performance of two full ring PET systems based on continuous crystals (cuboid and tapered geometries) coupled to SiPMs, focusing on sensitivity and spatial resolution enhancements.
研究成果
The study demonstrates that tapered continuous crystals offer a significant sensitivity enhancement (13.63% average gain) over cuboid crystals without degrading spatial resolution, using the selected position estimation method. The tighter packing of tapered crystals results in a more homogeneous sensitivity map and better recovery of homogeneous regions in image reconstruction.
研究不足
The results, especially spatial resolution, depend on the interaction position estimation method used and may not be generalizable to other methods. The study does not account for scatter from support structures, potentially underestimating scatter fraction.