研究目的
To introduce an ultrasensitive electrochemical aptamer-based sensor for detection of insulin in human blood serum using chitosan-carbon quantum dots as a new platform.
研究成果
The CQD-Chit nano-composite serves as an ideal material for the immobilization of insulin aptamer molecules and fabrication of a highly sensitive impedimetric insulin aptasensor. The aptasensor demonstrated high selectivity, good accuracy, low detection limit (106.8 pM), and high sensitivity in physiological pH, making it suitable for clinical applications.
研究不足
Potential saturation of the electrode surface at high insulin concentrations leading to deviation from linearity in the calibration curve.
1:Experimental Design and Method Selection:
Synthesis of carbon quantum dots (CQDs) from candle soot and dispersion in chitosan matrix to form CQD-Chit nano-composite. Characterization using TEM, FESEM, EDAX, FTIR, and electrochemical techniques.
2:Sample Selection and Data Sources:
Human blood serum samples for insulin detection.
3:List of Experimental Equipment and Materials:
Transmission electron microscopy (TEM), field emission scanning electron microscopy (FESEM), energy dispersive X-ray spectroscopy (EDAX), Fourier transform infrared (FTIR) spectroscopy, electrochemical impedance spectroscopy (EIS).
4:Experimental Procedures and Operational Workflow:
Fabrication of aptasensor by immobilizing insulin aptamer on CQD-Chit modified glassy carbon electrode. Detection of insulin using EIS.
5:Data Analysis Methods:
Analysis of Nyquist plots to determine charge transfer resistance (Rct) changes for insulin detection.