研究目的
To present the design, development, and characterization of a new graphite-walled cavity ionization chamber as a primary standard for air kerma rate for 137Cs and 60Co gamma radiation, including methods for cavity volume determination, correction factors, and re-evaluation according to ICRU90 recommendations.
研究成果
The new graphite-walled ionization chamber IGNAS-IC16A#001 was successfully characterized and shown to be reliable as a primary standard for air kerma rate, with results consistent with existing standards. It meets quality requirements and is recommended for adoption, though re-evaluation with ICRU90 recommendations indicates slight deviations that need consideration in future standards.
研究不足
The study is limited to gamma radiation from 137Cs and 60Co sources; other radiation types are not considered. The volume determination methods have uncertainties, and the weight method is affected by water surface tension. Chamber reassembly introduced variations in response, and environmental conditions like humidity require careful control. Monte Carlo simulations assume specific parameters and may not cover all real-world variations.
1:Experimental Design and Method Selection:
The study involved designing and constructing a graphite-walled cavity ionization chamber (IGNAS-IC16A#001) for use as a primary standard. Methods included geometric volume determination using coordinate measuring machines, weight methods, electric field analysis via finite-element methods, and Monte Carlo simulations for correction factors.
2:Sample Selection and Data Sources:
The chamber was tested with 137Cs and 60Co gamma radiation sources. Data on dimensions, ionization currents, and environmental conditions were collected.
3:List of Experimental Equipment and Materials:
Equipment included a coordinate measuring machine (SIP type CMM5), digital electrometer (Keithley model 6517A), thermometer (Elmetron model PT-401), barometer (Vaisala model PTB-200), hygrometer (Elmetron model PWT-401), and gamma radiation sources. Materials included ultra-pure graphite, aluminum, polyethylene heat-shrink tubing, and distilled water.
4:Experimental Procedures and Operational Workflow:
Procedures involved measuring chamber dimensions, determining cavity volume, conducting saturation, leakage current, stability, angular dependence, and polarity tests, and performing air kerma rate measurements with environmental corrections. Monte Carlo simulations were used to calculate physical constants and correction factors.
5:Data Analysis Methods:
Data were analyzed using statistical methods for uncertainty calculation, linear fitting for recombination correction, and Monte Carlo codes (EGSnrc) for simulation-based factors.