oe1(光电查) - 科学论文

2 条数据
?? 中文(中国)
  • Multiple renal ruptures after flexible ureteroscopic lithotripsy with holmium laser

    摘要: The authors present a case report of multiple renal ruptures after flexible ureteroscopic lithotripsy (FURL) with holmium laser. Multiple renal ruptures following flexible ureterorenoscopy have not been reported so far. The etiology remains unclear. We like to share this case to make urologists aware of this unusual complication and discuss possible causes and therapeutic approaches.

    关键词: urolithiasis,flexible ureteroscopy,infection,Renal rupture,diabetes mellitus

    更新于2025-09-19 17:13:59

  • Laser fibre, rather than the stone, may harm the scope: retrospective monocentric analysis of 26 pre- and intraoperative factors of flexible ureteroscope (fURS) damage

    摘要: Purpose The increasing number of flexible ureteroscopy procedures, the fragility of devices and their repair costs are a burden for urological departments worldwide. The objective was to investigate the impact of 26 pre- and intraoperative factors on reusable flexible ureteroscope (fURS) damage. Methods All procedures were conducted with reusable fURS: Karl Storz Flex-X2 or Olympus URF-V. Statistical analysis was performed in RStudio (1.0.136) with Chi-square test and Mann–Whitney U tests (MWU). Results In total, 416 flexible ureteroscopies, performed between September 2013 and June 2017, were analysed. 283 (68.03%) of these were for kidney stone surgery, and 133 (31.97%) for diagnostic purposes. In total, 39 (9.38%) devices were postoperatively deemed defective. The application of reusable laser fibre through fURS was more common in cases with documented defects [17/39 (43.59%) vs. 102/377 (27.06%), p = 0.047]. Other factors such as application of nitinol basket, biopsy via fURS, insertion of access sheath (UAS), as well as stone burden [median kidney stone maximal diameter: 6 mm (min 2.0; max 30.0) vs. 6 mm (min 1.0 vs. max 30.0)] showed no influence on fURS damage rate (p > 0.05). The infundibulopelvic angle (IPA) was steeper in cases with fURS damage as compared to cases without damage [median 44.0° (min 20.0; max 81.0) vs. 55.0 (min 7.0; max 122.0), p < 0.001]. Conclusions Application of laser fibre via fURS can be considered as a risk factor of fURS damage. Stone burden, as well as the usage of not-sharp ended devices as nitinol baskets or forceps, is primarily not responsible for fURS damage.

    关键词: Laser fibre,Flexible ureteroscopy,Infundibulopelvic angle,Damage

    更新于2025-09-11 14:15:04