研究目的
To further increase the sensitivity of DP-LIBS by proposing a concentric multipass cell (CMC) enhanced DP-LIBS (CMC-DP-LIBS) technique.
研究成果
The CMC-DP-LIBS technique significantly enhances the emission intensity and reduces the LOD of DP-LIBS, providing a new approach for trace element detection.
研究不足
The study mentions the potential for further optimization of CMC-DP-LIBS for higher enhancement, indicating current technical and application constraints.
1:Experimental Design and Method Selection:
The study employs a CMC-DP-LIBS setup to enhance the spectral signal and lower LOD by multi-reflection in CMC.
2:Sample Selection and Data Sources:
Seven samples for LOD determination of Mn were prepared by dropping different concentrations of MnCl2 solutions onto the surface of high purity zinc bulks.
3:List of Experimental Equipment and Materials:
Includes Nd:YAG lasers, digital delay generator, spectrometer, and optical fiber.
4:Experimental Procedures and Operational Workflow:
The setup involves an ablating laser and a reheating laser, with the latter's pulse reflected in the CMC for several times.
5:Data Analysis Methods:
The LOD was calculated according to the IUPAC method, and the signal enhancement factor was evaluated.