研究目的
To improve differentiation of polyolefins via AFM-IR by exploiting the measurement of viscoelastic properties to overcome the challenge of similar IR spectra.
研究成果
The study demonstrates that AFM-IR can be used to differentiate polymer domains with similar IR spectra by analyzing viscoelastic properties, specifically resonance frequency and decay time. This approach provides nanoscale resolution and improves contrast in material mapping. The correlation with DMTA data validates the sensitivity of AFM-IR to mechanical properties. Future work should focus on developing a robust theoretical framework for quantitative analysis.
研究不足
The technique is limited by the similarity in IR spectra of polyolefins, which complicates differentiation. The theoretical framework for quantitative viscoelastic analysis via AFM-IR is not fully developed, and the method may not differentiate materials with very similar mechanical properties without additional parameters like decay time. The AFM-IR measurements are sensitive to experimental conditions such as laser power and load force.
1:Experimental Design and Method Selection:
The study uses AFM-IR (Atomic Force Microscopy-Infrared spectroscopy) to combine morphological and compositional analysis. It employs a photothermally induced resonance (PTIR) technique to detect IR absorption with nanoscale resolution. A Kelvin-Voigt mechanical model is used to analyze viscoelastic properties from the cantilever deflection signals.
2:Sample Selection and Data Sources:
A copolymer of polypropylene (ICP) with ethylene-propylene rubber (EPR) inclusions is used as the sample, representing typical packaging film materials. Commercial polymers are also analyzed via dynamic mechanical thermal analysis (DMTA) for validation.
3:List of Experimental Equipment and Materials:
Equipment includes an AFM-IR setup (likely from Anasys Instruments, as implied by the tip reference), a ZnSe prism for IR reflection, and a DMTA setup. Materials include polypropylene (PP), ethylene-propylene rubber (EPR), and other polymers like EVOH, LDPE, nylon.
4:Experimental Procedures and Operational Workflow:
AFM-IR is performed with the cantilever in contact mode, scanning the sample surface. IR pulses at specific wavenumbers (e.g., 1460 cm?1) are applied, and the deflection signal of the cantilever is measured and Fourier transformed to extract resonance frequency and decay time. DMTA measurements are conducted to correlate mechanical properties.
5:Data Analysis Methods:
Data is analyzed by fitting the deflection signal to a damped harmonic oscillator model (Eq. 1) to obtain parameters like resonance frequency (ω) and decay time (τ). Fourier transformation is used to derive Lorentzian functions for further analysis. Correlation with DMTA-derived storage modulus (G') is performed.