研究目的
Investigating the role of magnesium dopants in LiF:Mg,Ti thermoluminescent dosimeters through electronic structure calculations to understand the defects induced by Mg and their impact on the dosimeter's efficiency.
研究成果
The study reveals that Mg dopants in LiF introduce defect states in the band-gap that can trap electrons, enhancing the dosimeter's efficiency. The most stable defects are those involving Mg interstitials and substitutions, which induce local lattice distortions. These findings provide a theoretical basis for the observed enhancement in thermoluminescent response of LiF:Mg,Ti dosimeters.
研究不足
The study relies on computational models and approximations, such as the choice of functional and supercell size, which may affect the accuracy of the results. Experimental validation of the predicted defect states and their energies is needed.
1:Experimental Design and Method Selection:
First-principles electronic structure calculations using density functional theory (DFT) with the CP2K code, employing Gaussian and plane waves (GPW) method. The PBE0 hybrid functional was used for accurate defect level calculations.
2:Sample Selection and Data Sources:
Calculations were performed on pristine LiF and LiF with various Mg-induced defects (interstitial Mg, substitutional Mg, Li vacancy plus Mg interstitial, etc.) using supercells containing 216 atoms.
3:List of Experimental Equipment and Materials:
Computational resources included the UK national high-performance computing (HPC) service, ARCHER, and UNAM HPC service, Miztli.
4:Experimental Procedures and Operational Workflow:
Structural optimizations and electronic structure calculations were performed with convergence criteria for forces and displacements. Spin-polarized calculations were carried out to account for unpaired electrons.
5:Data Analysis Methods:
Defect formation energies and thermodynamic transition levels were calculated to assess the stability and electronic properties of the defects. Projected density of states (PDOS) and spin density analyses were used to characterize the defect states.